INSTRUMENTS WITH A DIGITAL SOUL

Promotions

AX-3003P

Pwr sup.unit: programmable laboratory; Channels: 1; 0÷30VDC; 1mV

AX-CP-07-R

Clip-on probe; pincers type; 10A; red; Grip capac: max.4mm; 4mm

AX-5002

Temperature meter; LCD 4 digits, with a backlit; Resol: 0.1°C

AX-2040

AC digital clamp meter; LCD (4000), with a backlit; V DC: 4÷400V

AX-B150

Inspection camera; Display: LCD 2,4" (320x240); Cam.res: 640x480

AX-LCR42A

LCR meter; double LCD (19,999/1999), bargraph, with a backlit

AX-176

Digital multimeter; LCD (6600), with a backlit; 3x/s; True RMS

AX-T2090

Non-contact voltage and cable detector; LCD, bargraph

AX-CP-04-R

Clip-on probe; with puncturing point; 10A; red; 4mm

AX-7020

Analogue multimeter; Features: universal; V AC: 10/50/250/500V

AXIOMET Catalogue

Download Catalogue
(ver. 6)
PDF (12,9 MB)

What to do if an infrared thermometer does not allow for adjusting emissivity factor?

Emissivity factor value can be from 0 to 1.

In case of infrared thermometers which do not allow for adjusting this parameter manually, it is usually stored in internal memory of the device and it has a constant value of 0.95. It is suitable for most commonly examined materials.

A problem occurs in case of materials with a very smooth surface, such as polished metals, emissivity factor value of which is below 0.1.

In order to measure temperature of such objects, it would be good to buy an infrared thermometer which allows for entering emissivity factor manually. If it is not possible, there are some methods that will improve accuracy of temperature measurement:

  • Use insulation tapes specially designed for infrared thermometers; their emissivity factor is known. They can be put on the object under test. The insulation tape quickly absorbs the temperature of the object and its emissivity corresponds to the factor stored in internal memory of the infrared thermometer.
  • Determine errors of measurement on your own for different temperatures, within the range important for the user. When you compare readings from the infrared thermometer with readings from the tactile thermometer, you can further calculate measurements taken with the use of the infrared thermometer to obtain more credible values.

Both these methods can be used in case of surfaces with unknown emissivity factor.