INSTRUMENTS WITH A DIGITAL SOUL

Promotions

AX-EL600W

Electronic load; LCD 4,3", with a backlit; 0÷150V; 0.001÷60A

AX-LCR42A

LCR meter; double LCD (19,999/1999), bargraph, with a backlit

AX-DG100-SOF

Software; Application: AX-DG105; Equipment: RS232 cable, software

AX-LCR41A

LCR meter; LCD 5 digits; 3x/s; 0,0001÷9,999MΩ; R accuracy: ±0,3%

AX-C850

RTD calibrator; R range: 0,01÷400/1500/3200Ω

AX-C830

Thermocouple calibrator; V DC: -10m÷75mV; Measuring unit: °C, °F

AX-C708

Multimeter calibrator; V DC: 0,01m÷40m/400m/4/40/400V

AX-DG1005AF

Generator: function; Band: ≤5MHz; LCD TFT 3,5"; Channels: 2

AX-DG1015AF

Generator: function; Band: ≤15MHz; LCD TFT 3,5"; Channels: 2

AX-C605

Loop calibrator; V DC: 0÷28V; I DC: 0÷22mA; Current source: 0÷22mA

AXIOMET Catalogue

Download Catalogue
(ver. 6)
PDF (12,9 MB)

Product specifications
Type of meter:
digital multimeter
Measurement:
frequency, dBm, DC voltage, capacitance, DC current, resistance, duty cycle
Measuring instrument features:
  • auto and manual ranging
  • automatic power-off
  • HOLD function
  • MIN/MAX function
  • REL function
  • indicates incorrect connection of test leads
  • low battery indicator
Kind of display used:
LCD 5 digits
Bargraph:
51 segm.
Sampling:
2,5x/s
DC voltage measuring range:
0,001m...50m/500m/5m/50/500/1000V
DC voltage measuring accuracy:
±(0,03% + 10 digits)
AC/DC + AC voltage measurement:
0,001m...50m/500m/5V/50V/500V/1000V
AC/DC + AC voltage measuring accuracy:
±(0,5% + 40 digits)
Bandwidth for AC/DC + AC voltage measurement:
40...20 kHz
DC current measuring range:
0,01µ...500µ/5m/50m/500m/5/10A
DC current measuring accuracy:
±(0,15% + 15 digits)
AC/DC + AC current measurement:
0,01µ...500µ/5m/50m/500m/5A/10A
AC/DC + AC current measuring accuracy:
±(0,75% + 20 digits)
True effective value measurement:
True RMS
Bandwidth for AC/DC + AC current measurement:
40...20 kHz
Resistance measuring range:
0,01...500/5k/50k/500k/5M/50MΩ
Resistance measuring accuracy:
±(0,1% + 10 digits)
Capacitance measuring range:
0,01n...50n/500n/5µ/50µ/500µ/5mF
Capacitance measuring accuracy:
±(1% + 5 digits)
Frequency measuring range:
5...200kHz (linear waveform)
Frequency measuring accuracy:
±(0,006% + 4 digits)
Frequency measurement sensitivity:
100mV (for linear waveform), Vp 2-5V (for rectangular waveform)
Duty cycle factor:
0.01...95 %
Duty cycle factor measurement accuracy:
±(2% + 5 digits)
Waveform frequency for duty cycle factor:
5...500 Hz
Diode test:
0,7mA, 2,5V
Continuity test:
acoustic signal for R<50Ω
Pollution degree:
2
Overall dimensions (W x D x H):
100 x 40 x 200 mm
Weight with battery:
560 g
Power supply:
6 batteries 1,5V LR03 (AAA), set without battery
Interface:
  • IrDA
  • USB
Standard equipment:
  • set of test leads with probes
  • set of test leads with insulated clips
  • software (Windows 98, 2000, NT, XP, Vista)
  • optical cable with USB
Conform to the norm:
EN61010 1000V CAT II, EN61010 600V CAT III
Kind of output connector:
USB A plug
Protection:
anti-overload
Measuring resolution:
0,01Ω
Additional information
Gross weight:
0.911 kg